The Sierra Summit/Ascent system startup procedure is different than the Sierra Wave system start up. Powering up your Sierra Summit/Ascent system in the proper sequence is important for successful computer and equipment synchronization.
The proper sequence is:
- Turn on the Summit/Ascent computer and monitor first.
- Wait for Windows to completely start up until you see all of the desktop icons.
- Turn the Sierra Summit/Ascent base unit on.
- Wait several seconds for the Sierra equipment USB drivers to fully load.
- Start the Sierra Summit/Ascent program.
Note that the Sierra Wave startup sequence was different due to the different connection technology. Sierra Wave uses an Ethernet network connection, and the Sierra Summit/Ascent uses a USB connection to communicate with the computer.